Chromatography

Fast protein liquid chromatography

System Specifications

  • Control system – ChromLab Software
    (compatible across all NGC Systems)
  • Dimensions (W x D x H) – 49 x 61 x 56 cm (NGC Quest Plus and
    NGC Scout Plus Systems)
    49 x 61 x 74 cm (NGC Discover System)
  • Weight – 41–46 kg (NGC Quest Plus and
    (excluding computer) NGC Scout Plus Systems)
    64 kg (NGC Discover System)
  • Power supply – 100–240 V, 50–60 Hz
  • Power consumption – 750 W maximum

System Pump

  • Pump type – Reciprocating piston
  • Flow rate setting – 10 ml/min pumps: 0.001–10 ml/min
    (normal range)
    100 ml/min pumps: 0.01–100 ml/min
    (normal range)
  • Flow rate accuracy – ±2% (conditions: F10 pump — 0.1–10 ml/min,
    F100 pump — 1.0–100 ml/min;
    pressure: <600 psi [4.1 MPa, 41 bar];
    viscosity: 0.5–3.7 cP)
  • Pressure range – 10 ml/min pumps: 0–25.2 MPa (3,650 psi)
    100 ml/min pumps: 0–10 MPa (1,450 psi)
  • Viscosity range – 0.5–10.8 cP (for 10 ml/min and
    100 ml/min pumps)

Sample Pump

  • Pump type – Piston pump, metering type
  • Flow rate setting – 0.01–100 ml/min
  • Flow rate accuracy – ±2%
  • Pressure range – 0–10 MPa (1,450 psi)
  • Viscosity range – 0.5–10.8 cP

Mixer

  • Mixing principle – Chamber with magnetic stirrer
  • Mixer volume – 263 μl (included), 750 μl (included), 2 ml,
    5 ml (F10)
    750 μl (included), 2 ml (included), 5 ml,
    12 ml (F100)
  • Gradient composition – ±0.5% (conditions: 3–97%B,
  • accuracy – 0.25–10 ml/min F10 pumps)
    ±0.8% (conditions: 5–95%B,
    1–100 ml/min F100 pumps)

Valves

  • Type – Rotary valves and rocker solenoid
  • Number of valves – 1 inject valve, up to 2 x 8-port sample
    inlet valves, and 2 x 8-port buffer inlet,
    2 x 12-port outlet, and 3 x 5-port column
    switching valves
  • Functions – Loop selection
    (PEEK Loop and DynaLoop offerings)

Pressure Sensors

  • Placement of sensors Standard: after system pump
    Options: precolumn, postcolumn,
    sample pump
  • Range 0–3,650 psi
  • Accuracy ±2 psi or 2%, whichever is greater

Inlet Valves

  • Inlet A – 8 inlets
  • Inlet B – 8 inlets
  • Sample inlet – 8 or 15 inlets

Multi-Wavelength Detector II Module

  • Wavelength – 190–800 nm
    Up to 4 simultaneously
  • Absorbance range – 0–3.0 AU
  • Linearity – 0–2.5 AU within ±5%
  • Operating pressure – 700 psi (4.8 MPa)
  • Flow cells – Preparative: 2 mm (volume: 140 μl)
    Analytical: 5 mm (volume: 51 μl)
    Analytical: 10 mm (volume: 24 μl)
  • Conductivity reading range – 0.01–999.9 mS/cm
  • Accuracy – ±2%
  • Operating pressure – 0–700 psi (0–4.8 MPa)
  • Conductivity cell volume – Included in flow cell volume
  • Temperature – 4–60°C monitor range 
  • Temperature – ±2% monitor accuracy

pH Monitor

  • pH reading range – 0–14
  • Accuracy – ±0.1 pH unit within pH 2–12
  • Operating pressure – 0–70 psi with pH probe inline
    and 0–500 psi in bypass mode
  • Flow cell volume – 100 μl (210 μl including internal flow paths)

Pressure Sensors

  • Placement of sensors Standard: after system pump
    Options: precolumn, postcolumn,
    sample pump
  • Range 0–3,650 psi
  • Accuracy ±2 psi or 2%, whichever is greater

NGC Fraction Collector

  • Collection modes
    Collect All, Threshold, and Time/Volume windows
    Flow rate 0.01–200 ml/min
  • Collection rack options
    (each NGC Fraction Collector can accommodate 4 racks)
    96 x 13 mm tubes, 75 x 16 mm tubes,
    75 x 18 mm tubes, 27 x 50 ml tubes,
    2 x deep well microplates (24-/48-/96-well),
    96 x 1.5–2 ml capless tubes, 16 x 250 ml
    bottles, and 40 x unlimited volume
    prep-rack adaptors
  • Operating temperature  – 4–40°C
  • Dimensions –  (W x D x H) 42 x 60 x 54.5 cm

BioFrac Fraction Collector

  • Collection modes
    Time 0.02–99,999 min
    Volume 0.02–99,999 ml
    Flow rate 0.01–100 ml/min
  • Collection rack options 180 x 12–13 mm tubes, 120 x 15–16 mm
    tubes, 80 x 18–20 mm tubes, 168 x 1.5 ml
    microtubes, 24 x 30 mm tubes, 4 x 96-,
    48-, 24-, or 12-position microplates, 4 x
    250 ml bottles, and 20 x unlimited volume
    prep-rack adaptors
  • Operating temperature 4–40°C
  • Dimensions (W x D x H) 44.5 x 35.6 x 38.7 cm

Column Switching Valve

  • Five-column valve –  Can connect up to 5 columns with
    forward and reverse flow, column bypass
    capability, and column pressure readings

Buffer Blending Valve

  • Blending valve –  Standard in the NGC Scout Plus and
    NGC Discover Systems
    Automated inline buffer preparation from
    concentrated buffer and salt stocks
    Double the fluid output to 20 ml/min
    or 200 ml/min

Air Sensor Module

  • Number of sensors –  Up to 8 total air sensors (1 for end of
    sample detection, remaining are buffer)
  • Placement of End of buffer, end of sample
    built-in sensors –
  • Sensing principle –  Acoustic